Urgent Field Safety Notice: Medtronic Puritan Bennett 500 Series Ventilators - Length of Cumulative Use

According to the manufacturer, this Urgent Field Safety Notice involved a device in the United States or other countries.

WHAT IS THIS?

A Urgent Field Safety Notice is communication sent by a manufacturer to users or customers in relation to a corrective action taken by the manufacturer for technical or medical reasons to prevent or reduce the risk of a serious incident in relation to a device made available on the market.

EVENT
Type of EventUrgent Field Safety Notice
Notice ReferenceFA1380
Notice QtyNot Stated
Notice DateJune 2024
Notice LinkLink
Notes/DetailsAdditional test analyses suggest that cumulative use of over 14 years may potentially pose a harm to health. These conclusions are derived from chemical testing and risk assessment from literature, including animal laboratory testing. There are no conclusive human clinical data describing quantifiable risk. No customer complaints have been received reporting any of these potential harms.
ReasonAny single PB500 series ventilator should only be used for the 10-year labelled service life. If a patient requires ventilatory support for a longer period of time, cumulative sequential use of more than one PB500 series ventilator should be limited to 14 years. After 14 years of cumulative use, patients should be transitioned to an alternate device.
DEVICE
ManufacturerMedtronic
Model Names500 Series Ventilators
Model NumbersPB560, PB540, PB520
Part/Item/Catalog/
REF/Product Number
-
Device DescriptionVentilator
Serial Number(s)Not stated.
Lot #/Exp Date(s)N/A
Medtronic PB560 Ventilator
Medtronic PB560 Ventilator
CUSTOMER/PATIENT ACTIONS

Please complete and return the enclosed Customer Acknowledgement Form in the Notice as directed to confirm the receipt and understanding of this information.
• Please provide this notice to those who need to be aware within your organization or to any organization or person where the product has been transferred or distributed.
• Any single PB500 series ventilator should only be used for the 10-year labelled service life. If a patient requires ventilatory support for a longer period of time, cumulative sequential use of more than one PB500 series ventilator should be limited to 14 years. After 14 years of cumulative use, patients should be transitioned to an alternate device.
• When approaching 14 years of cumulative use of a PB500 series ventilator, it is recommended that patients or their caregivers discuss transitioning to an alternate ventilator with the prescribing physician.

CONTACT INFORMATION

If you have any questions regarding this notification, please contact your Medtronic sales representative.


PLEASE EMAIL OR FAX THIS ACKNOWLEDGEMENT TO:
nahar.s.alsurayi@medtronic.com

Medtronic PB560 Ventilator
Medtronic PB560 Ventilator
DISCLAIMER
Medical devices help to diagnose, prevent and treat many injuries and diseases. We are not suggesting or implying that any companies or other entities included in the Database engaged in unlawful conduct or otherwise acted improperly. The same device may have different names in different countries. This database is not intended to provide medical advice and patients should check with their doctors to determine if it contains relevant information and if such information has medical implications for them.

© 2024 Equiptrack

Recalls and Medical Equipment Maintenance Records for the Medical Equipment Industry including Consumers, Distributors and 3rd Party Resellers.

 Reports     |     Maintenance     |     Partners     |     About     |     Contact     |     Help     |     Log In    |   Sign Up

Equiptrack products and services are based only on information supplied to Equiptrack. Equiptrack does not have the complete history of every unique piece of equipment. Use the Equiptrack search as one important tool, along with an equipment inspection and functional check, to make a better decision about purchasing your next piece of used medical equipment.