Urgent Field Safety Notice: GE Healthcare Infinia and VG Series Nuclear Medicine Systems - Unintended Detector Motion

According to the manufacturer, this Urgent Field Safety Notice involved a device in the United States or other countries.

WHAT IS THIS?

A Urgent Field Safety Notice is communication sent by a manufacturer to users or customers in relation to a corrective action taken by the manufacturer for technical or medical reasons to prevent or reduce the risk of a serious incident in relation to a device made available on the market.

EVENT
Type of EventUrgent Field Safety Notice
Notice Reference 40908
Notice QtyAll systems not running on: V.25BK (Win2k), C.15BK or C.1523 (Windows XP), or V.60BK (Alll VG Operating Systems)
Notice DateJuly 2025
Notice LinkGE Infinia and VG Systems - Unintended Detector Motion
Notes/DetailsNone
ReasonGE HealthCare has become aware of a potential safety issue related to possible unintended radial detector motion, which can occur during patient setup or during a patient scan. This can occur if the system does not have the correct version of the gantry software installed. In addition to the incorrect software, several other mechanical failures would also have to be present in order for this issue to arise, including radial gear malfunctions, gear oil leaks, and brake pad misalignment. In the unlikely event that the above failures are all present at the same time and the unintended detector motion is not noticed by the technologist, the detector can slowly descend and contact the patient.
DEVICE
ManufacturerGE Healthcare
Model NamesInfinia (Infinia, Infinia Hawkeye, Infinia II, Infinia II Hawkeye 4 ), VG (Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH)
Model NumbersN/A
Part/Item/Catalog/
REF/Product Number
-
Device DescriptionThe system is intended to be used to perform general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and optional imaging features designed to enhance image quality in Oncology, Cardiology, Neurology and other clinical diagnostic imaging applications.
Serial Number(s)All Affected per the software revision.
Lot #/Exp Date(s)N/A
GE Healthcare Infinia and VG Series-Unintended Detector Motion
GE Healthcare Infinia and VG Series-Unintended Detector Motion
CUSTOMER/PATIENT ACTIONS

Check the gantry software version by following the instructions in Appendix A.


If the system has the correct gantry software version installed as indicated in Table 1,
you can continue to use your system.


If the software version is not one of the versions in Table 1, GE HealthCare
recommends that you discontinue use of your system and contact GE
HealthCare Service or your local Service Representative.

CONTACT INFORMATION

If you have any questions regarding this notification, please contact your GE Healthcare sales representative.

PLEASE EMAIL OR FAX THIS ACKNOWLEDGEMENT TO:
recall.40908@gehealthcare.com 

GE Healthcare Infinia and VG Series-Unintended Detector Motion
GE Healthcare Infinia and VG Series-Unintended Detector Motion
DISCLAIMER
Medical devices help to diagnose, prevent and treat many injuries and diseases. We are not suggesting or implying that any companies or other entities included in the Database engaged in unlawful conduct or otherwise acted improperly. The same device may have different names in different countries. This database is not intended to provide medical advice and patients should check with their doctors to determine if it contains relevant information and if such information has medical implications for them.

© 2024 Equiptrack

Recalls and Medical Equipment Maintenance Records for the Medical Equipment Industry including Consumers, Distributors and 3rd Party Resellers.

 Reports     |     Maintenance     |     Partners     |     About     |     Contact     |     Help     |     Log In    |   Sign Up

Equiptrack products and services are based only on information supplied to Equiptrack. Equiptrack does not have the complete history of every unique piece of equipment. Use the Equiptrack search as one important tool, along with an equipment inspection and functional check, to make a better decision about purchasing your next piece of used medical equipment.